Cyclic voltammetry with the Rodeostat

electrochemistry potentiostat rodeostat

In a previous blog post we explored the irreversible redox reaction of ascorbate using the Rodeostat potentiostat. In this blog post we carried out a cyclic voltammetry experiment with ruthenium hexamine Ru(NH3)6 a reversible redox couple. 

An example voltammogram of this redox couple is demonstrated in the Pine electrode literature and we have previously described this experiment using the Cheapstat potentiostat. Here we have repeated this cyclic voltammetry experiment using the Rodeostat potentiostat. A Pine screen printed carbon electrode was also used. 

Equipment

Reagents

1) 0.1M KCl 

To prepare 500 mL of a 0.1M KCl solution: transfer 33 mL of 3M KCl (LabChem Part # LC18795-1) into a 500 mL volumetric flask. Fill to line with distilled or deionized water. Place stopper in flask and invert several times to mix.

2) 10 mM Ruthenium hexamine 

Hexaammineruthenium(III) chloride is available from Sigma-Aldrich (Cat # 262005). A 250 mg sample size is plenty of material for most experiments. Prepare a stock solution of 10 mM Ru(NH3)6 in 0.1M KCl as follows:

  • Add 4 mL of 0.1M KCl to the bottle containing 250 mg of 10mM Ru(NH3)6. Mix until dissolved. This is a 0.2 M stock solution.
  • Add 50 µL of the 0.2 M stock solution to 1 mL of 0.1 M KCl in a microcentrifuge tube.

      Cyclic voltammetry parameters

      • Start: -500 mV
      • Stop: 300 mV
      • Slope: 50 mV/sec
      • Sample rate: 3 mV/sample
      • Cycles: 3

      50 µL of sample was pipetted onto the Pine electrode. After taking several background measurements with 0.1M KCl, the ruthenium hexamine was tested. Samples were also run with the Cheapstat potentiostat for comparison.

      Results

      Fig 1. The result shown below is data collected from the 3rd cycle of the CV run with the Rodeostat potentiostat. 

      Fig. 2. Data collected from the 2nd and 3rd cycles of the CV run with the Rodeostat and Cheapstat potentiostats. You can see that the two CV runs from the Rodeostat (shown in red) overlap and are hard to distinguish. Like the data from the ascorbate experiments, there is less background noise in the data collected with the Rodeostat likely due to the higher resolution analog-to-digital converter and precision components.   



      Older Post Newer Post


      • Ernestreink on

        Металлоконструкции на заказ

      • Rogertes on

        https://xn—116-qdd9albfz5a2cwb.xn—p1ai/%D1%88%D0%B8%D0%BD%D0%BE%D0%BC%D0%BE%D0%BD%D1%82%D0%B0%D0%B6/

      • Ratnavali Arts on

        Wao it’s a great blog, Thank You! for sharing this information.
        Ratnavali Arts ( Best Jewellery Wholesaler in Jaipur, Jewellery shop in Jaipur )
        https://www.ratnavaliarts.com/
        Jaipur jewellery wholesale online, jewellery wholesaler in jaipur

      • yesdone samjho ho gaya on

        It’s a great blog, Thank You! for sharing this information. I will love to read your upcoming blogs.
        Yes Done ( Carpenter Service in Jaipur, Carpenter Repair Work, Carpenter in Jaipur )
        https://yesdone.in/services/Carpenter
        Carpenter for repair, Professional carpenter, Carpentry repair services

      • TBP Spice Exporter on

        “This blog is really informative, Thank You! for sharing this information. I have also bookmarked your blog.

        Best Spice Exporter
        https://bestspiceexporter.com/

        best spice exporter in India, top spice exporter in India, organic spices manufacturers in India"



      Leave a comment